Portfolio and departmental competencies
Blood Gas Analyser for Point of Care Testing
Cobas b123 training for clinical medical staff only
This online learning is also compatible with mobile devices.
This e-learning course will provide an overview of the Cobas b123 blood gas analyser, and will refresh your knowledge on the following topics to ensure that you are able to obtain reliable blood gas results:
•Running sample procedures
•Troubleshooting
•Support
This training will be followed by a multiple choice online assessment – you will need to pass this to retain access (user barcode) to the analyser.

Correct Use of Point of Care Testing Devices
Before proceeding with this training course, please click on the following link and read the Correct Use of Point of Care Testing Devices statement.
By continuing with this training course you are confirming that you have read, understood and accept the Trust’s policy relating to the use of your personal Point of Care Testing barcode and the use of correct patient identification on Point of Care Testing devices.
You are also confirming that you will abide by the Point of Care Testing policy.
Introduction to the Cobas b123 System
The Cobas b 123 POC system is a multi-parameter cartridge-based blood gas analyser. The following parameters can be measured:
•Blood gases (pO2, pCO2, pH)
•Electrolytes (Sodium, Potassium, Calcium)
•Metabolites (Glucose, Lactate)
•Haemoglobin derivatives (Methaemoglobin, Carboxyhaemoglobin)
•Total haemoglobin
•Oxygen saturation
Introduction to the OVERVIEW screen




Introduction to the WORKPLACE screen

Introduction to the INSTRUMENT screen

Introduction to the UTILITIES screen

Parameters
A parameter can have different colours, depending on the status. There are 5 possible statuses:





-
Green: the parameter is activated and ready.
-
Light green: the parameter is temporarily disabled
-
Red: the parameter is not ready because it is not calibrated or remotely locked.
-
Red with ampoule: the parameter is not ready due to a QC lock.
-
Transparent: the parameter is permanently deactivated
When a parameter is not ready (red), you can see the click on the [Info] button to find out what the problem is.
Test Procedure & Results
1. Pre-analytical considerations
Transport: Syringe samples are stable at room temperature for a maximum of 15 minutes, or 30 minutes if refrigerated. Capillary samples must be analysed immediately.
Mixing: Mix the sample by rolling the syringe between the hands or gently inverting. DO NOT shake the sample as this will cause the blood cells to fracture, leading to inaccurate results.
Immediately prior to measurement the sample should be gently mixed for at least 15 seconds to ensure that there has been no settling of red blood cells

Expel the first blood drop: Always expel the first drop of blood to remove any contaminants
Sample identification: It is important to ensure that samples are clearly labelled and that the patient ID on the samples matches that on the accompanying ICE request form.
2. Operator identification
In order to use the analyser you have to press the [Log on] button and scan or type in your barcode number. Enter the same for both Operator ID and Password.

The user ID barcode issued following training is for your personal use only. It is against Trust policy to share your barcode with any other staff member.
3. Running a sample
In order to run a sample the system must be ‘Ready’. Ensure the correct parameters are selected – these can be selected/deselected manually or you can use the parameter panel buttons.

As an example, to start a capillary measurement you must press [Start capillary measurement]. Then the system is preparing the sample input for a capillary sample.

1. Introduce the capillary gently to the inlet until it is held in place and confirm on-screen by pressing the [Yes] button.
